A polarimeter [1] is a scientific instrument used to measure the angle of rotation caused by passing polarized light through an optically active substance.[2]

Some chemical substances are optically active, and polarized (uni-directional) light will rotate either to the left (counter-clockwise) or right (clockwise) when passed through these substances. The amount by which the light is rotated is known as the angle of rotation. The angle of rotation is basically known as observed angle.

History

Polarization by reflection was discovered in 1808 by Étienne-Louis Malus (1775–1812).[2] (See also Optical rotation#History). A nicol prism or a polaroid filter can be used to polarize the light.

Measuring Principle

The ratio, the purity, and the concentration of two enantiomers can be measured via polarimetry. Enantiomers are characterized by their property to rotate the plane of linear polarized light. Therefore, those compounds are called optically active and their property is referred to as optical rotation. Light sources such as a light bulb, a light-emitting diode (LED), or the sun emit electromagnetic waves at the frequency of visible light. Their electric field oscillates in all possible planes relative to their direction of propagation. In contrast to that, the waves of linear-polarized light oscillate in parallel planes.[3]

If light encounters a polarizer, only the part of the light that oscillates in the defined plane of the polarizer may pass through. That plane is called the plane of polarization. The plane of polarization is turned by optically active compounds. According to the direction in which the light is rotated, the enantiomer is referred to as dextro-rotatory or levo-rotatory.

The optical activity of enantiomers is additive. If different enantiomers exist together in one solution, their optical activity adds up. That is why racemates are optically inactive, as they nullify their clockwise and counter clockwise optical activities.

The optical rotation is proportional to the concentration of the optically active substances in solution. Polarimeters may therefore be applied for concentration measurements of enantiomer-pure samples. With a known concentration of a sample, polarimeters may also be applied to determine the specific rotation when characterizing a new substance.

The specific rotation is a physical property and defined as the optical rotation α at a path length l of 1 dm, a concentration c of 1g/100 mL, a temperature T (usually 20 °C) and a light

wavelength λ (usually sodium D line at 589.3 nm):

[4]{{rp|123}}

This tells us how much the plane of polarization is rotated when the ray of light passes through a specific amount of optically active molecules of a sample. Therefore, the optical

rotation depends on temperature, concentration, wavelength, path length, and the substance

being analyzed.

[5]

Construction

The polarimeter is made up of two Nicol prisms (the polarizer and analyzer). The polarizer is fixed and the analyzer can be rotated. The prisms may be thought of as slits S1 and S2. The light waves may be considered to correspond to waves in the string. The polarizer S1 allows only those light waves which move in a single plane. This causes the light to become plane polarized.

When the analyzer is also placed in a similar position it allows the light waves coming from the polarizer to pass through it. When it is rotated through the right angle no waves can pass through the right angle and the field appears to be dark. If now a glass tube containing an optically active solution is placed between the polarizer and analyzer the light now rotates through the

plane of polarization through a certain angle, the analyzer will have to be rotated in same angle.

Operation

Polarimeters measure this by passing monochromatic light through the first of two polarising plates, creating a polarized beam. This first plate is known as the polarizer.[6] This beam is then rotated as it passes through the sample. After passing through the sample, a second polarizer, known as the analyzer, rotates either via manual rotation or automatic detection of the angle. When the analyzer is rotated such that all the light or no light can pass through, then one can find the angle of rotation which is equal to the angle by which the analyser was rotated(θ) in the former case or (90-θ) in the latter case.

Types of polarimeter

Laurent's half-shade polarimeter

When plane-polarised light passes through some crystals, the velocity of left-polarised light is different from that of the right-polarised light thus the crystals are said to have two refractive indices i.e. double refracting

Construction: It consists of a monochromatic source S which is placed at focal point of a convex lens L. Just after the convex lens there is a Nicol Prism P which acts as a polariser. H is a half shade device which divides the field of polarised light emerging out of the Nicol P into two halves generally of unequal brightness. T is a glass tube in which optically active solution is filled. The light after passing through T is allowed to fall on the analyzing Nicol A which can be rotated about the axis of the tube. The rotation of analyser can be measured with the help of a scale C. Laurent’s half-shade polarimeter is shown in Figure 6.22.

Working: In order to understand the need of a half-shade device, let us suppose that half-shade device is not present. The position of the analyzer is so adjusted that the field of view is dark when tube is empty. The position of the analyzer is noted on circular scale. Now the tube is filled with optically active solution and it is set in its proper position. The optically active solution rotates the plane of polarization of the light emerging out of the polariser P by some angle. So the light is transmitted by analyzer A and the field of view of telescope becomes bright. Now the analyzer is rotated by a finite angle so that the field of view of telescope again become dark. This will happen only when the analyzer is rotated by the same angle by which plane of polarization of light is rotated by optically active solution.

The position of analyzer is again noted. The difference of the two readings will give you angle of rotation of plane of polarization (8).

A difficulty is faced in the above procedure that when analyzer is rotated for the total darkness, then it is attained gradually and hence it is difficult to find the exact position correctly for which complete darkness is obtained.

To overcome above difficulty half-shade device is introduced between polariser P and

glass tube T.

Half Shade Device

It consist of two semicircular plates ACB and ADB. One half ACB is made of glass while other half is made of quartz.

Both the halves are cemented together. The quartz is cut parallel to the optic axis. Thickness of the quartz is selected in such a way that it

introduces a path difference of ’A/2 between ordinary and extraordinary ray. The thickness of the glass is selected in such a way that it absorbs the same amount of light as is absorbed by quartz half.

Let us consider that the vibration of polarisation is along OP. On passing through the glass half the vibrations remain along OP. But on passing through quartz half these vibrations will split into 0- and £-components. The £-components are parallel to the optic axis while O- component is perpendicular to optic axis. The O-component travels faster in quartz and hence an emergence 0-component will be along OD instead of along OC. Thus components OA and OD will combine to form a resultant vibration along OQ which makes same angle with optic axis as OP. Now if the Principal plane of the analyzing Nicol is parallel to OP then the light will pass through glass half unobstructed. Hence glass half will be brighter than quartz half or we can say that glass half will be bright and the quartz half will be dark. Similarly if principal plane of analyzing Nicol is parallel to OQ then quartz half will be bright and glass half will be dark.

When the principal plane of analyzer is along AOB then both halves will be equally bright. On the other hand, if the principal plane of analyzer is along DOC. then both the halves will be equally dark.

Thus it is clear that if the analyzing Nicol is slightly disturbed from DOC then one half becomes brighter than the other. Hence by using half shade device, one can measure angle of rotation more accurately.

Determination of Specific Rotation

In order to determine specific rotation of an optically active substance (say sugar) the polarimeter tube T is first filled with pure water and analyzer is adjusted for equal darkness (Both the halves should be equally dark) point. The position of the analyzer is noted with the help of scale. Now the polarimeter tube is filled with sugar solution of known concentration and again the analyser is adjusted in such a way that again equally dark point is achieved. The position of the analyzer is again noted. The difference of the two readings will give you angle of rotation θ. Hence specific rotation S is determined by using the relation.

[S]t λ =θ /LC

The above procedure may be repeated for different concentration.

Biquartz polarimeter

In biquartz polarimeters, a biquartz plate is used. Biquartz plate consists of two semi circular plates of quartz each of thickness 3.75mm. One half consists of right-handed optically active quartz,while the other is left-handed optically active quartz.

Lippich polarimeter

X-Ray Polarimeter

Quartz-Wedge polarimeter

Manual

The earliest polarimeters, which date back to the 1830s, required the user to physically rotate one polarizing element (the analyzer) whilst viewing through another static element (the detector). The detector was positioned at the opposite end of a tube containing the optically active sample, and the user used his/her eye to judge the "alignment" when least light was observed. The angle of rotation was then read from a simple fixed to the moving polariser to within a degree or so.

Although most manual polarimeters produced today still adopt this basic principle, the many developments applied to the original opto-mechanical design over the years have significantly improved measurement performance. The introduction of a half-wave plate increased "distinction sensitivity", whilst a precision glass scale with vernier drum facilitated the final reading to within ca. ±0.05º. Most modern manual polarimeters also incorporate a long-life yellow LED in place of the more costly sodium arc lamp as a light source.

Semi-automatic

Today, semi-automatic polarimeters are available. The operator views the image via a digital display adjusts the analyzer angle with electronic controls.

Fully automatic

Fully automatic polarimeters are now available and simply require the user to press a button

and wait for a digital readout. Fast automatic digital polarimeters yield an accurate result within a second, regardless of the rotation angle of the sample. In addition, they provide continuous measurement, facilitating High-performance liquid chromatography and other kinetic investigations.

Another feature of modern polarimeters is the Faraday modulator. The Faraday modulator creates an alternating current magnetic field. It oscillates the plane of polarization to enhance the detection accuracy by allowing the point of maximal darkness to be passed through again and again and thus be determined with even more accuracy.

As the temperature of the sample has a significant influence on the optical rotation of the sample, modern polarimeters have already included Peltier Elements to actively control the temperature. Special techniques like a temperature controlled sample tube reduce measuring errors and ease operation. Results can directly be transferred to computers or networks for automatic processing.[7]

Traditionally, accurate filling of the sample cell had to be checked outside the instrument, as an appropriate control from within the device was not possible. Nowadays a camera system allows accurate monitoring of the sample and filling conditions in the sample cell from inside the instrument.

A telecentric camera gives a sharp image over the complete length of any sample cell placed within modern instruments. The online monitoring of the filling process ensures that no bubbles or particles obstruct the measurement. A picture can be saved together with the recorded data. Any temperature gradients, inhomogeneous sample distributions or air bubbles can immediately be recognized before measurement, so that potential errors caused by bubbles or particles are no longer an issue.

Sources of error

The angle of rotation of an optically active substance can be affected by:

  • Concentration of the sample
  • Wavelength of light passing through the sample (generally, angle of rotation and wavelength tend to be inversely proportional)
  • Temperature of the sample (generally the two are directly proportional)
  • Length of the sample cell (input by the user into most automatic polarimeters to ensure better accuracy)
  • Filling conditions (bubbles, temperature and concentration gradients)

Most modern polarimeters have methods for compensating or/and controlling these errors.

Calibration

Polarimeters can be calibrated – or at least verified – by measuring a quartz plate, which is constructed to always read at a certain angle of optical rotation (usually +34°, but +17° and +8.5° are also popular, depending on the sample). Quartz plates are preferred by many users because solid samples are much less affected by variations in temperature, and do not need to be mixed on-demand like sucrose solutions.[8]

Applications

Because many optically active chemicals such as tartaric acid, are stereoisomers, a polarimeter can be used to identify which isomer is present in a sample – if it rotates polarized light to the left, it is a levo-isomer, and to the right, a dextro-isomer. It can also be used to measure the ratio of enantiomers in solutions.

The optical rotation is proportional to the concentration of the optically active substances in solution. Polarimetry may therefore be applied for concentration measurements of enantiomer-pure samples. With a known concentration of a sample, polarimetry may also be applied to determine the specific rotation (a physical property) when characterizing a new substance.

Chemical industry

Many chemicals exhibit a specific rotation as a unique property (an intensive property like refractive index or Specific gravity) which can be used to distinguish it. Polarimeters can identify unknown samples based on this if other variables such as concentration and length of sample cell length are controlled or at least known. This is used in the chemical industry.

By the same token, if the specific rotation of a sample is already known, then the concentration and/or purity of a solution containing it can be calculated.

Most automatic polarimeters make this calculation automatically, given input on variables from the user.

Food, beverage and pharmaceutical industries

Concentration and purity measurements are especially important to determine product or ingredient quality in the food & beverage and pharmaceutical industries. Samples that display specific rotations that can be calculated for purity with a polarimeter include:

{{div col|colwidth=30em}}
  • Steroids
  • Diuretics
  • Antibiotics
  • Narcotics
  • Vitamins
  • Analgesics
  • Amino acids
  • Essential oils
  • Polymers
  • Starches[9]
  • Sugars
{{div col end}}

Polarimeters are used in the sugar industry for determining quality of both juice from sugar cane and the refined sucrose. Often, the sugar refineries use a modified polarimeter with a flow cell (and used in conjunction with a refractometer) called a saccharimeter. These instruments use the International Sugar Scale, as defined by the International Commission for Uniform Methods of Sugar Analysis (ICUMSA).

See also

{{commons category|Polarimeters}}
  • Optical rotation
  • Polarimetry
  • Polarization
  • Chirality
  • Enantiomers

References

1. ^polarimeter. Princeton WordNet
2. ^Polarimeter. kenyon.edu
3. ^Hart, C. (2002), Organische Chemie, Wiley-VCH, {{ISBN|3-527-30379-0}}
4. ^{{cite book |title= Advanced Organic Chemistry, Part A: Structure and Mechanisms |edition= Fifth |publisher= Springer |date= 2007 |author= F. A. Carey |author2= R. J. Sundberg |DOI= 10.1007/978-0-387-44899-2}}
5. ^{{cite web|url=http://www.anton-paar.com/corp-en/products/group/polarimeter/|title=Group :: Anton-Paar.com|website=Anton Paar}}
6. ^Polarimetry {{webarchive|url=https://web.archive.org/web/20110927160932/http://www.files.chem.vt.edu/chem-ed/spec/molec/polarim.html |date=2011-09-27 }}. chem.vt.edu
7. ^Krüss Optronic "Polarimeters", Version 1.0, Hamburg, March 2012
8. ^Quartz Plate Calibration Standard. Rudolph Research
9. ^{{cite web|url=http://www.anton-paar.com/?eID=documentsDownload&document=31259&L=0|title=Starch|publisher=}}